Hydrogen cyanide

NO.12D
HCN
Performance
Measuring Range 1 to 200 ppm
Sampling Hours 1 to 10 hours
Detecting Limit 0.4 ppm (10 hours)
Colour Change Yellow → Red
Reaction Principle 2HCN + HgCl2 → Hg(CN)2 + 2HCl
HCl + Base → Chloride
Coefficient of Variation 10% (for 10 to 200 ppm-hr)
Shelf Life 3 Years
Corrections for temperature & humidity Humidity correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Ammonia   No
Nitrogen dioxide   + Red
Hydrogen sulphide   + Red
Amines   No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
- C 4.7 ppm 5.6 to 40%


Return to previous viewGo to top of page