Trichloroethylene |
NO.132LL |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1.5 minutes per pump stroke | |||
| Detecting Limit | 0.05 ppm (n=2) | |||
| Colour Change | Yellow → Purple | |||
| Reaction Principle | Cl2C:CHCl + PbO2 + H2SO4 → HCl HCl + Basic Compound → Chloride |
|||
| Coefficient of Variation | 10% (for 0.25 to 1 ppm), 5% (for 1 to 4 ppm) | |||
| Shelf Life | 2 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Hydrogen chloride, Chlorine, Bromine | 1/2 time | + | Pale bluish purple |
| Tetrachloroethylene | 1/2 time | + | Pale bluish purple |
| 1,1,1-Trichloroethane | 80 ppm | No | No ( 80 ppm) |
| Toluene, Xylene | No | No |
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| 1,2-Dichloroethylene | 1.5 | 1 | 0.375 to 6 ppm |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 10 ppm | 25 ppm | - |