Tetrachloroethylene |
NO.133L |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1.5 minutes per pump stroke | |||
| Detecting Limit | 0.4 ppm (n=2) | |||
| Colour Change | Yellow → Pink | |||
| Reaction Principle | Cl2C:CCl2 + PbO2 + H2SO4 → HCl HCl + Base → Chloride |
|||
| Coefficient of Variation | 10% (for 2 to 5 ppm), 5% (for 5 to 25 ppm) | |||
| Shelf Life | 2 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Nitric Oxide, Nitrogen dioxide | No | No | |
| Hydrogen chloride, Chlorine, Bromine | + | Pink | |
| Acetone | 200 ppm | No | No |
| Unsaturated Halogenated HCs | + | Pink | |
| Aromatic hydrocarbons | 100 ppm | No |
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Pentachloroethane | 20 | 1 | 40 to 500 ppm |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 25 ppm | 100 ppm | - |