| Measuring Range |
| 6 to 20 ppm |
20 to 200 ppm |
200 to 900 ppm |
|
| Number of Pump Strokes |
|
| Correction Factor |
|
| Sampling Time |
3 minutes per pump stroke |
| Detecting Limit |
3 ppm (n=2) |
| Colour Change |
White → Pale Pink |
| Reaction Principle |
1,1,1-Trichloroethane reacts with oxidizing agent to produce intermediate products then it reacts with detecting agent to produce pale pink stain. |
| Coefficient of Variation |
10% (for 20 to 60 ppm), 5% (for 60 to 200 ppm) |
| Shelf Life |
2 Years |
| Corrections for temperature & humidity |
Temperature correction is necessary |
| Store the tubes at cool and dark place. |