Methyl bromide |
NO.136LL |
![]() |
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 2 minutes per pump stroke | ||
| Detecting Limit | 0.02 ppm (n=2) | ||
| Colour Change | White → Pale purple | ||
| Reaction Principle | CH3Br + CrO3 + H2S2O7 → Br2 Br2 +3,3'-DIMETHYLNAPHTHIDINE → reaction product |
||
| Coefficient of Variation | 10% (for 0.1 to 0.4 ppm), 5% (for 0.4 to 1.2 ppm) | ||
| Shelf Life | 2 Years | ||
| Corrections for temperature & humidity | Temperature correction is necessary | ||
| Store the tubes at cool and dark place. | |||
| Substance | Concentration | Interference | Change colour by itself to |
| Chlorine | + | Pale purple( |
|
| Trichloroethylene | + | Pale purple( |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 1 ppm | - | 10 to 15% |