Carbon monoxide |
NO.1LM |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 45 seconds per pump stroke | |||
| Detecting Limit | 5 ppm (n=2) | |||
| Colour Change | White → Pale brown / Pale green (dual layers) | |||
| Reaction Principle | 5CO + I2O5 + H2S2O7 → I2 | |||
| Coefficient of Variation | 10% (for 50 to 200 ppm), 5% (for 200 to 1000 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Acetylene, Olephines, Propane, Hexane | + | Pale brown | |
| Butane | + | Pale brown | |
| Hydrogen | No | No |
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 25 ppm | - | 12.5 to 74% |