Chlorine Dioxide |
NO.23M |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.05 ppm (n=5) | |||
| Colour Change | White → Pale pink | |||
| Reaction Principle | Chlorine dioxide oxidises the indicator to produce Pale pink discolouration. | |||
| Coefficient of Variation | 10% (for 0.05 to 0.2 ppm), 5% (for 0.2 to 0.6 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes in dark and cool place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Nitric oxide | + | Pale pink | |
| Ozone | + | Pale pink | |
| Nitrogen dioxide | + | Pale pink | |
| Chlorine | + | Pale pink | |
| Bromine, Iodine | + | Pale pink |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.1 ppm | 0.3 ppm | - |