Carbon dioxide |
NO.2HH |
![]() |
| Measuring Range |
|
||
| Number of Pump Strokes |
|
||
| Correction Factor |
|
||
| Sampling Time | 45 seconds | ||
| Detecting Limit | 0.04% (n=1) | ||
| Colour Change | Orange → Yellow | ||
| Reaction Principle | CO2 + N2H4 → NH2 |
||
| Coefficient of Variation | 10% (for 5 to 10 %), 5% (for 10 to 40 %) | ||
| Shelf Life | 3 Years | ||
| Corrections for temperature & humidity | Unnecessary | ||
| Store the tubes at cool and dark place. | |||
| Substance | Concentration | Interference | Change colour by itself to |
| Ammonia | No | No | |
| Carbon monoxide | No | No | |
| Hydrogen chloride | No | White ( |
|
| Chlorine | No | White ( |
|
| Nitrogen dioxides | No | Pale yellow ( |
|
| Sulphur dioxide | No | Pale yellow ( |
|
| Ethanol | No | Pale yellow ( |
|
| Acetic acid | No | Pale yellow ( |
|
| Hydrogen sulphide | No | Pale yellow ( |
|
| Ttimethylamine | No | ||
| Nitoric oxide | No | ||
| Ethylene , Propane | No |
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 5000 ppm | 30000 ppm | - |