Hydrogen sulphide

NO.4HM
H2S
Performance
Measuring Range
25 to 50 ppm 50 to 800 ppm 800 to 1600 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2
Sampling Time 1 minute per pump stroke
Detecting Limit 5 ppm (n=2)
Colour Change White → Brown
Reaction Principle H2S + Pb(CH3COO)2 → PbS + 2CH3COOH
Coefficient of Variation 5% (for 50 to 800 ppm)
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
กก กก กก กก
Other substance measurable with this detector tube
Substance Correction Factor No. of pump strokes Measuring Range
Nitrogen dioxide 1/4 time   No
sulphur dioxide 1/1 + No
Ammonia   No No
Carbon dioxide   No No
Carbon monoxide   No No
Mercaptans   No No
Organic gases & vapours   No No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
1 ppm 5 ppm 4.0 to 44%