Water vapour |
NO.6 |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1 minute per pump stroke | |||
| Detecting Limit | 0.1 mg/L (n=2) | |||
| Colour Change | Yellowish Green → Purple | |||
| Reaction Principle | H2O + Mg(ClO4)2 → Mg(ClO4)2 |
|||
| Coefficient of Variation | 10% (for 1 to 6 mg/L), 5% (for 6 to 18 mg/L) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Alcohols, Ketones, Aldehydes, sulphur dioxide, Hydrogen chloride | 2000 ppm | + | No |
| Esters, Acrylonitrile | 5000 ppm | + | No |
| Ammonia | 1000 ppm | + | No |
| Nitrogen dioxide, Hydrogen cyanide |
500 ppm | + | No |
| Hydrogen sulphide | No | No |
| Calibration gas generation | Vaper pressure method |
| TLV-TWA | TLV-STEL | Explosive range |
| - | - | - |