Phosphine |
NO.7 |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1.5 minutes per pump stroke | |||
| Detecting Limit | 0.5 ppm (n = 4) | |||
| Colour Change | White → Pale yellow | |||
| Reaction Principle | 2PH3 + 6HgCl2 + 3H2O → Hg3P2 |
|||
| Coefficient of Variation | 10% (for 5 to 10 ppm), 5% (for 10 to 50 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Arsine | 20 ppm | + | Brown |
| Hydrogen selenide | 20 ppm | + | Brown |
| Hydrogen sulphide | 10 ppm | + | Dark brown |
| Calibration gas generation | High pressure gas cylinder method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.3 ppm | 1 ppm | - |