Ethyl mercaptan |
NO.72L |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 1.5 minutes per pump stroke | |||
| Detecting Limit | 0.15 ppm (n=2) | |||
| Colour Change | Yellow → Red | |||
| Reaction Principle | C2H5SH + HgCl2 →ClH5S |
|||
| Coefficient of Variation | 10% (for 0.5 to 5 ppm), 5% (for 5 to 30 ppm) | |||
| Shelf Life | 2 Years | |||
| Corrections for temperature & humidity | Temperature correction is necessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Hydrogen sulphide | + | Red | |
| Mercaptans | + | Red | |
| Acetylene | + | No | |
| Carbon monoxide | + | No | |
| Ethylene | + | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 10 ppm | - | 2.8 to 18.2% |