Chlorine |
NO.8H |
![]() |
| Measuring Range |
|
|||
| Number of Pump Strokes |
|
|||
| Correction Factor |
|
|||
| Sampling Time | 45 seconds per pump stroke | |||
| Detecting Limit | 5 ppm (n=2) | |||
| Colour Change | White → Vermilion (reddish orange) | |||
| Reaction Principle | Cl2 + o - Tolidine → Vermilion products | |||
| Coefficient of Variation | 10% (for 50 to 100 ppm), 5% (for 100 to 500 ppm) | |||
| Shelf Life | 3 Years | |||
| Corrections for temperature & humidity | Unnecessary | |||
| Store the tubes at cool and dark place. | ||||
| Substance | Concentration | Interference | Change colour by itself to |
| Bromine | + | Vermilion | |
| Chlorine dioxide | + | Vermilion | |
| Iodine | + | Vermilion |
| Substance | Correction Factor | No. of pump strokes | Measuring Range |
| Chlorine dioxide | 0.5 | 1 | 25 to 250 ppm |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 0.5 ppm | 1 ppm | - |