Sulphide

NO.201H
 
Performance
Measuring Range 0.02 to 0.2 mg
Sampling Time 45 seconds per 1 pomp stroke
Detecting Limit 0.002 mg
Colour Change White → Brown
Reaction Principle Gus generation tube : Sulphides + H2SO4 → H2S
Detector tube : H2S + Pb(CH3COO)2 → PbS
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Coefficient of Variation 5% (for 0.02 to 0.2 mg)
 
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Changes colour by itself to
Chloride ion   No No
Chromate ion   No No
Carbonate ion   No No
Calibration gas generation Static gas dilution method


Return to previous viewGo to top of page