sulphide Ion

NO.211
S2-
Performance
Measuring Range 1 to 100 ppm
Sampling Time 2 minutes
Detecting Limit 0.5 ppm
Colour Change White → Brown
Reaction Principle S2- + Pb(CH3COO)2 → PbS
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Changes colour by itself to
CO32- 500 ppm + No
Cl- 5000 ppm + No
CrO42- 25 ppm Yellow
SO32- 300 ppm No
S2O32- 50 ppm Unclear demarcation No


Return to previous viewGo to top of page