sulphide Ion

NO.211LL
S2-
Performance
Measuring Range 0.5 to 20 ppm
Sampling Time 2 minutes
Detecting Limit 0.2 ppm
Colour Change White → Brown
Reaction Principle S2- + Pb(CH3COO)2 → PbS
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
SO32- 20 ppm No No ( 1%)
Cl- 100 ppm No No ( 5%)
Cr6+ 0.2 ppm Yellow ( 1ppm)
CO32- 500 ppm No No ( 1%)
S2O32- 20 ppm No No ( 1%)
I- 200 ppm No Yellow ( 3000ppm)
SO42- 100 ppm No No( 1%)
PO43- 100 ppm No No ( 1%)