Chloride Ion

NO.221L
Cl-
Performance
Measuring Range (25) to 1000 mg/L
Sampling Time 3 minutes
Detecting Limit 1 mg/L
Colour Change Brown → White
Reaction Principle 2Cl- + Ag2CrO4 → 2AgCl
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
CN- 1 mg/L + White
Br- 10 mg/L + White
SCN- 30 mg/L + White
Fe2+ 200 mg/L No ( 1000 mg/L)
Fe3+ 200 mg/L White
F- 500 mg/L No ( 1000 mg/L)
I- 50 mg/L + White
S2- 5 mg/L + Blackish gray
SO42- 2000 mg/L White


Return to previous viewGo to top of page