Chloride Ion

NO.221LL
Cl-
Performance
Measuring Range 10 to 200 mg/L
Sampling Time 1 minutes
Detecting Limit 3 mg/L
Colour Change Brown → White
Reaction Principle 2Cl- + Ag2CrO4 → 2AgCl + CrO42-
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Changes colour by itself to
CN- 10 mg/L + White ( 1mg/L)
Br- 10 mg/L + White ( 10mg/L)
SCN- 1 mg/L + White ( 100mg/L)
Fe2+ 20 mg/L No ( 1000mg/L)
Fe3+ 20 mg/L + White ( 20mg/L)
F- 1000 mg/L No No( 1000mg/L)
I- 20 mg/L + Pale yellow ( 20mg/L)
S2- 3 mg/L + Blackish gray ( 3mg/L)
SO42- 1000 mg/L No No ( 1000mg/L)


Return to previous viewGo to top of page