Performance
Measuring Range (5) to 50 mg/L
Sampling Time 5 minutes
Detecting Limit 1.0 mg/L
Colour Change White → Orange
Reaction Principle Fe2+ + 1,10 Fenantoroline (White) → Complex compound
Shelf Life 3 Years
Corrections for temperature & humidity Unnecessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Zn 5 mg/L + No ( 100mg/L)
Co 5 mg/L + Orange ( 1mg/L)
CN- 20 mg/ L No ( 100mg/L)
Fe3+ 50 mg/L + Pale orange ( 50 mg/L)
CU+ 5 mg/L Pale orange ( 1 mg/L)
CU2+ 10 mg/L + Pale blue ( 50 mg/L)
Ni 5 mg/L + No ( 100 mg/L)
Mn 50 mg/L No ( 100 mg/L )
PO43- 100 mg/L No No ( 100 mg/L)