Zinc Ion

NO.285
Zn
Performance
Measuring Range 3 to 20 mg/L
Sampling Time 3 minutes
Detecting Limit 0.5 mg/L
Colour Change Pale orange → Reddish purple
Reaction Principle Zn + Indicator → Reddish purple complex compound
Shelf Life 3 Years
Corrections for temperature & humidity Water temperature correction is necessary
Store the tubes at cool and dark place.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
CN- 10 mg/L + No ( 100mg/L)
Fe2+ 1 mg/L + Reddish purple ( 3mg/L)
Fe3+ 2 mg/ L + No ( 100mg/L)
Ni 2 mg/L + Purple(whole layer) ( 2mg/L)
Cu 0.1 mg/L + Reddish purple ( 0.1mg/L)
Mn 0.5 mg/L + Purple ( 3mg/L)
Pb 0.5 mg/L + No ( 100mg/L)


Return to previous viewGo to top of page