| GSP-250FT Specifications | |
| Air Pump: | Diaphragm type |
| Real-time Flow Rate Parameters: | 10 to 250 mL/min (optional range of 251 to 300 mL/min) |
| Calibration to Zero Flow Rate: | Internal components provide auto-correction for zero flow rate |
| Display: | digital display |
| Real-time Flow Rate: | mL/min minimum value on scale: 1mL |
| Integrated Flow Rate: | 0.00 to 9.99L minimum value on scale: 0.01L 10.0 to 99.9L minimum value on scale: 0.1L 100 to 999L minimum value on scale: 1L |
| Intake Time Display: | minute increments |
| minimum value on scale: | 1.0 min |
| Accuracy of Real-time / Integrated Flow Rate values: | standard deviation +/-5% (for a minimum real-time flow rate of 80 mL/min at fixed temperature) |
| Continuous Operation Time: | Approx. 30 hours when fully recharged |
| Standard Operating Temperature Range: | 5 to 35ºC(-5 to 40ºC also possible) |
| Power Source: | 7.2V Ni-cad battery and AC power (for adaptor) |
| Dimensions: | 142W x 99D x 172H (mm) |
| Weight: | 2.8kg |

| GSP-2LFT Specifications | |
| Air Pump: | Diaphragm type |
| Real-time Flow Rate Parameters: | 0.1 to 2.0 L/min |
| Calibration to Zero Flow Rate: | Internal components provide auto-correction for zero flow rate |
| Display: | Liquid crystal digital display |
| Real-time Flow Rate: | L/min minimum value on scale: 0.01L |
| Integrated Flow Rate | 0.01 to 99.9L minimum value on scale: 0.01L 100 to 999L minimum value on scale: 0.1L 1000 to 9999L minimum value on scale: 1L |
| Intake Time Display: | Minute increments |
| Minimum value on scale: | 1min |
| Accuracy of Real-time / Integrated Flow Rate values: | standard deviation +/-5% (for a minimum real-time flow rate of 0.5 L/min at fixed temperature) |
| Continuous Operating Time: | Approx. 8 hours at full power |
| Standard operating temperature range: | 5-35ºCi5-40ºC can effectively be usedj |
| Power Source: | 7.2V Ni-Cad battery and AC power (for adaptor) |
| Dimensions: | 160W x 135D x 183H (mm) |
| Weight: | 3.8kg |
|
Collection Tube Heater
[GTH-1]
|
||
The air sampler collects aldehydes which are hazardous air pollutants. The ozone scrubber is connected to the front stage and two DNPH collection tubes to the rear for air sampling. However, excessive moisture due to rain may interfere with collection by causing condensation in the collection tube. The GTH-1 prevents condensation by heating the collection tube to a temperature higher than the dew point with a direct current voltage (10ºC to 15ºC higher than the ambient air temperature) thus enabling reliable sampling on a 24-hour basis.
|
|
|
|
Branch Tube
  [GSP-2BRA]
![]() Max. flow volume: 40mL/min (2 x 20mL/min) |
Holder for Collection Tubes
[GSP-SH]
![]() |
All-Weather Cover
  [RC-1]
![]() |
Carrying Case
  [GSP-K]
![]() |
|||
|
Gas Sampling Pump Tripod
[U-8000]
![]() Extension length: 590mm Height: 1,500mm Weight: 1,500g |
|