Hexane |
NO.102TP |
![]() |
| This detector tube is only to be used with automatic gas sampler GSP-300FT-2 |
| Calibration Range | 2 - 80 ppm |
| Measuring range | 2 - 80 ppm |
| Sampling (flow) Rate | 100mL/min |
| Sampling Volume | 1000mL |
| Sampling Time | 10min. |
| Detecting Limit | 0.5 ppm |
| Colour Change | Orange → Dark green |
| Corrections for temperature/humidity | Unnecessary for 0 to 40ºC Unnecessary for R.H.0 to 90% |
| Shelf life | 3 years |
| Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
| Substance(s) | Interference | Colour change by itself |
| Sulphur dioxide | + | Dark green | |
| Hydrogen sulphide | + | Dark green | |
| Acetylene | + | Dark green | |
| Alcohols | No | No | |
| Esters, Ketones | + | Dark green | |
| Organic solvents(C3 or higher) | + | Dark green | |
| Aromatic hydrocarbons | + | Dark green |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 50ppm | -- | 1.1¡Á7.5% |