Methanol |
NO.111TP |
![]() |
| This detector tube is only to be used with automatic gas sampler GSP-300FT-2 |
| Calibration Range | 20 - 300 ppm |
| Measuring range | 20 - 300 ppm |
| Sampling (flow) Rate | 50mL/min |
| Sampling Volume | 500mL |
| Sampling Time | 10min |
| Detecting Limit | 5 ppm |
| Colour Change | Pink → Pale blue |
| Corrections for temperature/humidity | Necessary for 0 to 40ºC Unnecessary for R.H.10 to 90% |
| Shelf life | 2 years |
| Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
| Substance(s) | Interference | Colour change by itself |
| Alcohols | + | Pale blue | |
| Esters, Ketones | No | No | |
| Aliphatic hydrocarbons | No | No | |
| Aromatic hydrocarbons | No | No |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 200 ppm | 250 ppm | 6 to 36% |