Methanol

NO.111TP
CH3OH


This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range 20 - 300 ppm
Measuring range 20 - 300 ppm
Sampling (flow) Rate 50mL/min
Sampling Volume 500mL
Sampling Time 10min
Detecting Limit 5 ppm
Colour Change Pink → Pale blue
Corrections for temperature/humidity Necessary for 0 to 40ºC
Unnecessary for R.H.10 to 90%
Shelf life 2 years
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Alcohols   + Pale blue
Esters, Ketones   No No
Aliphatic hydrocarbons   No No
Aromatic hydrocarbons   No No
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
200 ppm 250 ppm 6 to 36%