Isopropyl alcohol

NO.113TP
CH3CH¡ÊOH¡ËCH3¡Êi-C3H7OH¡Ë


This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range 20 - 200 ppm
Measuring range 20 - 400 ppm
Sampling (flow) Rate 100mL/min
Sampling Volume 1000mL,500mL
Sampling Time 10min. 5min.
Detecting Limit 5 ppm
Colour Change Pink → Pale blue
Corrections for temperature/humidity Necessary for 0 to 40ºC
Unnecessary for R.H.10 to 90%
Shelf life 2 years
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Alcohols   + Pale blue
Esters, Ketones   No No
Aliphatic hydrocarbons   No No
Aromatic hydrocarbons   No No
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
200 ppm 400 ppm 2 to 12.7%