Xylene |
NO.123TP |
![]() |
This detector tube is only to be used with automatic gas sampler GSP-300FT-2 |
| Calibration Range | 2 - 80 ppm |
| Measuring range | 2 - 80 ppm |
| Sampling (flow) Rate | 100mL/min |
| Sampling Volume | 1000mL |
| Sampling Time | 10min |
| Detecting Limit | 1 ppm |
| Colour Change | White → Brown |
| Corrections for temperature/humidity | Unnecessary for 0 to 40ºC Unnecessary for R.H.0 to 90% |
| Shelf life | 2 years |
| Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
| Substance(s) | Interference | Colour change by itself |
| Carbon monoxide | Unclear demarcation | Brown (When 50ppm or higher) | |
| Ethyl benzene | + | Brown | |
| Ethyl acetate | 400ppm |
No | No |
| Dichloro methane | 30ppm |
No | No |
| N,N,dimethylformamide | 25ppm |
No | No |
| Trichloroethylene | + | No | |
| Toluene | + | Brown | |
| n-Hexane | Unclear demarcation | No | |
| Methanol | 200ppm |
No | No |
| Calibration gas generation | Diffusion tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 100 ppm | 150 ppm | 1.0¡Á7.0% |