Vinyl chloride

NO.131P
CH2¡§CHC£ì



This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range 50 - 1500 μg/m3
Measuring range 50 - 1500 μg/m3
Sampling (flow) Rate 100mL/min
Sampling Volume 3000mL
Sampling Time 30min
Detecting Limit 10μg/m3
Colour Change Yellow → Pale reddish purple
Corrections for temperature/humidity Necessary for 5 to 35 ºC
Unnecessary for R.H.20 to 80%
Shelf life 2 years
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Ammonia   No No
Hydrogen chloride,Chlorine   No No
Nitrogen oxides   No No
p-Dichlorobenzene   + Pale reddish purple
1,2-Dichloroethylene   + Pale reddish purple
Trichloroethylene   + Pale reddish purple
Tetrachloroethylene   + Pale reddish purple
1,1,1-Trichloroethane   No No
Aromatic hydrocarbons   No No
Formaldehyde   No No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
1 ppm -- 3.6 to 23%