Ethylene oxide

NO.163TP
C2H4O
This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range¡§ 0.1-5 ppm
Measuring range 0.1-5 ppm
Sampling (flow) Rate 50mL/min
Sampling Volume 500mL
Sampling Time 10 min.
Detecting Limit 0.05 ppm(500mL)
Colour Change yellow → Pale orange
Corrections for temperature/humidity Necessary for 0 to 40ºC
Unnecessary for R.H.10 to 90%
Shelf life 1 year (in the refrigerator)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Acetaldehyde 2.0 ppm + Pale orange
Ketones 15 ppm + Pale orange
Alcohols 300 ppm ¡¼ No
Formaldehyde 0.6 ppm + Pale orange
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
1ppm ¡¼¡¼ 3.0¡Á100%