Chlorine

NO.8TP
Cl2


This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range 0.05 - 0.6 ppm
Measuring range 0.05 - 0.6 ppm
Sampling (flow) Rate 100mL/min
Sampling Volume 1000mL
Sampling Time 10min.
Detecting Limit 0.02 ppm
Colour Change Pink → White
Corrections for temperature/humidity Unnecessary for 0 to 40ºC
Unnecessary for R.H.10 to 90%
Shelf life 2 year
Accuracy CV=5-10%
(CV: Coefficient of variation = σ: standard deviation / mean value x 100)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Ammonia ≧1/6 No
Sulphur dioxide   No
Nitrogen dioxide   No No
Hydrogen sulphide   No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
0.5ppm 1 ppm ーー