Formaldehyde

NO.91TP
HCHO



This detector tube is only to be used with automatic gas sampler GSP-300FT-2
Specifications
Calibration Range (0.01) - 0.50 ppm
Measuring range 0.01 - 1.75 ppm
Sampling (flow) Rate 100mL/min , 50mL/min
Sampling Volume 1000mL , 500mL
Sampling Time 10min
Detecting Limit 0.005ppm (1000mL)
Colour Change Yellow → Pale orange
Corrections for temperature/humidity Temperature correction is necessary
Shelf life 1 year (in the refrigerator)
Possibly coexisting substances and their interferences (NOTE)
Substance(s) Interference Colour change by itself
Acetaldehyde   + Pale orange
Acetone   + Pale orange
Ethanol 0.15% Yellow
Xylene 50ppm No No
Ammonia 1ppm Yellow
Nitrogen dioxide 1ppm No No
Calibration gas generation Permeation tube method
TLV-TWA TLV-STEL Explosive range
--

C0.3ppm

7.0 - 73%