Nitrogen dioxide |
NO.9P |
![]() |
| This detector tube is only to be used with automatic gas sampler GSP-300FT-2 |
| Calibration Range | 0.02 - 0.20 ppm |
| Measuring range | 0.02 - 0.20 ppm |
| Sampling (flow) Rate | 100mL/min |
| Sampling Volume | 3000mL |
| Sampling Time | 30min |
| Detecting Limit | 0.004 ppm |
| Colour Change | White → Orange brown |
| Corrections for temperature/humidity | Necessary for 0 to 40ºC Unnecessary for R.H.0 to 90% |
| Shelf life | 2 years |
| Accuracy | CV=5-10% (CV: Coefficient of variation = σ: standard deviation / mean value x 100) |
| Substance(s) | Interference | Colour change by itself |
| Carbon monoxide | No | No | |
| Nitric oxide | No | No | |
| Chlorine | 0.10ppm |
No | Reddish brown from 0.08ppm |
| Ozone | 0.04ppm |
No | Pale yellow from 0.1ppm |
| Carbon dioxide | No | No | |
| Acetone | No | No | |
| p-Dichlorobenzene | No | No | |
| n-decan | No | No | |
| Aromatic hydrocarbons | No | No | |
| Formaldehyde | No | No |
| Calibration gas generation | Permeation tube method |
| TLV-TWA | TLV-STEL | Explosive range |
| 3 ppm | 5 ppm | -- |