Pull the pump handle
Point the tube towards the target area and simply wait until sampling has been completed.
|
The number of sampling strokes and the sampling time
The number of sampling strokes and the sampling time differs for the individual detector tube types.
Be sure to always consult the respective instruction sheet or users manual.
The following procedures should be used to ensure that the correct volume has been aspirated.
1.Wait until the sampling time has elapsed. Completion of 100mL or 50mL sampling can be verified using the flow finish indicator in the handle.
|
|
2. Turn the handle 90 degrees and let go; it should remain in place without retracting.
Otherwise, repeat the procedure (of pulling out the handle) until it remains in place without retracting.
|
|
 |
Reading the colour change
After sampling (the time allowed to measure the presence of target gas) has been completed, the colour change layer in the calibrated tube is read to determine the correct concentration.
|
Tips for detector tube reading
 |
When the end of the colour change layer is flat, simply read the value at the end. |
 |
When the end of the colour change layer is slanted, read the value in the middle of the slant. |
 |
When the demarcation of the colour change layer is pale, the mean value between the dark and the pale layer ends is taken. |
 |
 |
 |
 |
 |
 |
 |
 |
 |
In this case, the reading would be 5%. |
 |
In this exaggerated case, the reading would be 5%. |
 |
In this exaggerated case, the reading would be 5%. |
 |
|
Tip for easier reading
| When you marked the colour change with a pen as soon as the sampling was completed, it is more convenient to read. |
|
|