GASTEC CORPORATION - For all types of gas and vapour

Automatic Air Sampling PumpGSP-501FT

Compact, lightweight, and reliable automatic air sampling pump for GASTEC detector tubes and sorbent tubes.

  • Flow range 50 - 500mL/min
  • 10 - 500mL/min is available when volume mode (When set to 10 - 49 mL/min, the intermittent operation is performed at 50mL/min)
  • Fixed flow function automatically regulates load changes
  • Flow rate, sampling time, and integrated volume are displayed on the same screen simultaneously
  • Quiet operation with low noise
  • Automatic start enables sampling to begin after a preset time
  • 5 sampling settings can be saved with the program mode
  • Real-time flow volume and integrated flow volume are automatically corrected to the values for 20ºC(68ºF) or 25ºC(77ºF)
Automatic Air Sampling Pump GSP-501FT

Specifications

Sampling mode Timer Mode:Air pump automatically stops at set time Settable time: 1 minute to 30 hours
Volume Mode:Air pump automatically stops at set volume Settable integrated volume: 0.010 to 900L
Settable instantaneous flow rate Timer Mode:50-500mL/min
Volume Mode:10-500mL/min
(When set to 10-49 mL/min, the intermittent operation is performed at 50mL/min)
Constant flow rate
operating range
10-49mL/min:0.0-5.0kPa  300mL/min:0.0-23.0kPa
50mL/min:0.0-40.0kPa     400mL/min:0.0-16.0kPa
100mL/min:0.0-37.0kPa   500mL/min:0.0-10.0kPa
200mL/min:0.0-30.0kPa
Display Liquid crystal digital display (with backlight), Display range:0-600mL/min
Structure and function Constant flow rate function (built-in set flow rate holding circuit),
Autostart function (autostart after set standby time in standby mode),
Diaphragm type air pump, Program Mode (5 sampling settings)
Accuracy of instantaneous flow measurement Instantaneous flow rate: 50-500 mL/min ±5%
Accuracy of integrated flow
Measurement
[When set flow rate: 50 to 500mL/min] ± 5%.
<Volume Mode only>
[When set flow rate: 10 to 49mL/min] ±(2.5×sampling time [min]) mL.
Operating temperature range 0-40℃
Operating humidity range 10-90% RH (non-condensing)
Power supply 2 AA alkaline batteries (standard accessories, commercially available)
2 AA nickel-metal hydride batteries (commercially available)
Continuous operation time 2 AA alkaline batteries (Standard Accessories): 20 hours
(Set flow rate: 200 mL/min, suction pressure: 2kPa or less, ambient temperature: 25°C)
Dimensions and weight 80(W) ×40(D)×126(H) mm 280g (including batteries)
Standard accessories 2 AA alkaline batteries , detector tube adaptor, tube tip holder, dust filter (5 pcs), instruction manual, warranty certificate, inspection certificate
Directives and regulations EU Directive:2014/30/EU(EMC), 2011/65/EU,(EU)2015/863(RoHS)
UK Regulation :2016 No. 1091(EMC),2012 No.3032(RoHS)
EMC harmonised standards EU:EN 61326-1:2013
UK:BS EN 61326-1:2013
RoHS designated standards EU:EN IEC63000:2018
UK:BS EN IEC63000:2018

Intermittent Operation
When the instantaneous flow rate is set to 10 - 49mL/min in the Volume Mode, the intermittent operation is performed by sampling at an instantaneous flow rate of 50mL/min.

 
Example of Intermittent Operation
When the Volume Mode is selected and the flow rate is set to 10 mL/min and the set volume is 0.1L, after the sampling is started, the pump will pause for about 48 seconds after about 12 seconds of sampling at a flow rate of 50mL/min (samples until the sampled volume per minute reaches 10mL, then pauses). Sampling at a flow rate of 50 mL/min for about 12 seconds and pausing for about 48 seconds are repeated until the set volume of 0.1 L is reached, taking about 10 minutes.

Sorbent tubes

Product name/code Layer Filling quantity Dimensions(mm) Tubes/box Shelf life (months)
Activated charcoal tube 251S-20 2 100/50 5.6×100 20 60
251S2-20 1 150 5.6×100 20 60
Activated charcoal tube (Bead-shaped) 258-20 2 100/50 5.6×100 20 60
258A-20 2 400/200 7.0×105 20 60
258S2-20 1 150 5.6×100 20 60
Silica-gel tube 252S-20 2 400/200 7.0×105 20 60
252S2-20 1 600 7.0×105 20 60
252S3-20 2 150/75 5.6×100 20 60
252S4-20 1 300 5.6×100 20 60

Detector tube

Substance to be measured Tube no. & name Measuring range
(ppm)
Scale range
(ppm)
No. of
tests
per box
Detector tube type Notes
Acetone 151TP Acetone 25-800 (25)-800 10 Automatic air sampling pump
Acrylonitrile 191TP Acrylonitrile 0.2-12.6 0.2-3.0 5 Automatic air sampling pump
Ammonia 3S Ammonia 0.5-5 0.5-5 10 Automatic air sampling pump Sampling of hazardous air pollutants
Benzene 121TP Benzene 0.1-14.5 0.1-5 5 Automatic air sampling pump
121P Benzene 250-3000µg/m3 (250)-3000µg/m3 5 Automatic air sampling pump Sampling of hazardous air pollutants
Carbon disulphide 13TP Carbon disulphide 0.05-2.4 (0.05)-1.00 5 Automatic air sampling pump
Chlorine 8TP Chlorine 0.05-0.6 0.05-0.6 10 Automatic air sampling pump Residential environment
p-Dichlorobenzene 127P p-Dichlorobenzene 100-3000µg/m3 100-3000µg/m3 5 Automatic air sampling pump Residential environment
N,N-Dimethylacetamide 183TP N,N-Dimethylformamide 3.0-57.5 10 Automatic air sampling pump
N,N-Dimethylformamide 183TP N,N-Dimethylformamide 0.5-30 (0.5)-15 10 Automatic air sampling pump
Ethyl benzene 122P Toluene 110-2750µg/m3 5 Automatic air sampling pump Residential environment
Ethylene oxide 163TPM Ethylene oxide 1-50 1-50 5 Automatic air sampling pump
163TP Ethylene oxide 0.1-5 0.1-5 5 Automatic air sampling pump
Formaldehyde 91P Formaldehyde 0.02-1.44 0.02-0.4 10 Automatic air sampling pump Residential environment Indoor environment
91PL Formaldehyde 0.01-0.80 (0.01)-0.20 10 Automatic air sampling pump Residential environment Indoor environment
91TP Formaldehyde 0.01-1.75 0.01-0.50 10 Automatic air sampling pump
Hexane 102TP Hexane 2-80 2-80 10 Automatic air sampling pump
Hydrogen cyanide 12TP Hydrogen cyanide 0.3-9.0 0.3-4.5 10 Automatic air sampling pump
Hydrogen fluoride 17TP Hydrogen fluoride 0.05-9.0 0.05-3.0 10 Automatic air sampling pump
Hydrogen sulphide 4S Hydrogen sulphide 10-200ppb 10-200ppb 5 Automatic air sampling pump Sampling of hazardous air pollutants
4TP Hydrogen sulphide 0.1-2.88 0.1-1.6 10 Automatic air sampling pump
Isopropyl alcohol 113TP Isopropyl alcohol 20-400 20-200 10 Automatic air sampling pump
Methanol 111TP Methanol 20-300 20-300 10 Automatic air sampling pump
Methyl ethyl ketone 152TP Methyl ethyl ketone 20-300 20-300 10 Automatic air sampling pump
Nitrogen dioxide 9P Nitrogen dioxide 0.02-0.20 0.02-0.20 10 Automatic air sampling pump Residential environment
Styrene 124S Styrene 0.2-4.0 0.2-4.0 5 Automatic air sampling pump Sampling of hazardous air pollutants
Tetrachloroethylene 133TP Tetrachloroethylene 2.5-84 2.5-40 5 Automatic air sampling pump
133P Tetrachloroethylene 20-720µg/m3 20-300µg/m3 5 Automatic air sampling pump Sampling of hazardous air pollutants
Toluene 122TP Toluene 2-80 2-80 5 Automatic air sampling pump
122P Toluene 100-7000µg/m3 100-2500µg/m3 5 Automatic air sampling pump Residential environment
Trichloroethylene 132TP Trichloroethylene 1-33 1-15 5 Automatic air sampling pump
132P Trichloroethylene 20-1200µg/m3 20-500µg/m3 5 Automatic air sampling pump Sampling of hazardous air pollutants
Vinyl chloride 131TP Vinyl chloride 0.2-9.6 0.2-3.0 10 Automatic air sampling pump
131P Vinyl chloride 50-1500µg/m3 50-1500µg/m3 5 Automatic air sampling pump Sampling of hazardous air pollutants
Xylene 123TP Xylene 2-80 2-80 5 Automatic air sampling pump
122P Toluene 540-13500µg/m3 5 Automatic air sampling pump Residential environment

How to use

1.
Break tips off the detector tube using the tube tip holder.
2.
Insert the detector tube to the inlet of the rubber tube holder.
Put the power switch to ON. Adjust the flow rate, confirm that the timer of the air sampling pump is set, then push the start button.
3.
After sampling, remove the detector tube from the rubber tube holder and read the concentration of the tube.

Options

Mount GSP-501FT to use at the desired height.

Case for personal exposure measurement

Tube holder for personal exposure measurement

Enables the detector tube to be securely set up for measuring and helps to avoid possible injury to the user as well as damage to the detector tube itself.

Reusable rubber caps can be attached to the broken tip of the detector tubes as an extra precautionary safety measure. One (1) package contains 20 caps.

2 pumps can be mounted on the same tripod